FAQ for DXAdvanced Series

DXAdvanced SeriesPaperless Recorder

  • Display DX100P/DX200P, DX1000/DX2000 (/AS1 model) settings on the Universal Viewer

    Install DAQSTANDARD (model DXA120, 1 license comes standard with the DX1000/DX2000) in the same folder as the SMARTDAC+ STANDARD Universal Viewer. When you install DAQSTANDARD, select “Custom” for the Setup type, and change the installation destination folder to the SMARTDAC+ STANDARD Universa...

    Publishd:
    Mar, 2016
    ID:
    w0000078
    Category:
    Software
  • Inputting settings on the DAQSTANDARD (DXA120) hardware configurator software

    Enter "ˆC." The caret (ˆ) symbol is on the "6" key. (Use only a single-byte character.)

    Publishd:
    Jan, 2018
    ID:
    w0001141
    Category:
    Setting
  • Error code "code flash sum error."

    The internal flash memory is experiencing an error. Contact your dealer or Yokogawa representative for repairs.

    Publishd:
    Nov, 2017
    ID:
    w0001130
    Category:
    Other
  • Can I use DAQSTANDARD on Windows 7?

    DAQSTANDARD supports Windows 7 version R8.22.01 or later. If you have an unsupported version, download and install the latest version. For the license number, use the universal key.

    Publishd:
    Nov, 2017
    ID:
    w0001131
    Category:
    Specification
  • If I set the DX1000/DX2000 to High Speed mode, does the number of input channels decrease?

    No, even in High Speed mode, the number of input channels is the same as Normal mode. Note: In High Speed mode, the integral time of the A/D transformer is fixed at 600 Hz, and the power-line period noise rejection performance will be lowered. If measuring in an environment susceptible to power-li...

    Publishd:
    Nov, 2017
    ID:
    w0001132
    Category:
    Specification
  • E0401 communication error.

    Check the following items. Whether the LAN cable is correctly connected, and whether the PC's wireless LAN function is OFF. Whether the IP addresses of the recorder and PC are set. Whether you are using the latest version of DAQSTANDARD. Mode is Key+Comm. Whether the same ...

    Publishd:
    Nov, 2017
    ID:
    w0001133
    Category:
    Other
  • Can I use DAQSTANDARD on Windows 10?

    Starting with version R9.03.06, DAQSTANDARD supports Windows 10. If you have an unsupported version, download and install the latest version. During installation, use license number 116-00000-8807 when prompted.

    Publishd:
    Nov, 2017
    ID:
    w0001134
    Category:
    Software
  • Upper/lower limit of scale in the DX1000/DX2000 input range settings.

    In the DX1000/DX2000 range settings, the allowable setting range from lower limit to upper limit is -30000 to 30000 (excluding the decimal point). Change the upper or lower limit value or the decimal point to stay within that range. The decimal place is determined by the scale lower limit. For exam...

    Publishd:
    Nov, 2017
    ID:
    w0001135
    Category:
    Setting
  • Can channels be added to the DX1000/DX2000?

    With the DX1000/DX2000, you cannot add channels.

    Publishd:
    Nov, 2017
    ID:
    w0001136
    Category:
    Specification
  • DX1000/DX2000 integral flow settings.

    (1) As an example, enter the following formulas for calculating flow on MATH channel 1 and MATH channel 2. 101=TLOG.SUM(001) 102=TLOG.SUM(001) (2) MATH channel 1 resets hourly, and MATH channel 2 resets daily. TLOG rolling average, turn reset of each MATH channel ON/OFF, and specify a ...

    Publishd:
    Nov, 2017
    ID:
    w0001137
    Category:
    Setting
  • Preventing upper-limit alarms during measurement of temporarily rising temperature.

    Yes, set a delay upper-limit alarm. A delay upper-limit alarm will not engage until the measured value remains above a specified value for a specified length of time (the delay time).

    Publishd:
    Nov, 2017
    ID:
    w0001138
    Category:
    Setting
  • Monitoring the DX1000/DX2000 screen on a PC.

    You can monitor the DX1000/DX2000 screen from Internet Explorer. Use the following steps. Set the IP address on the recorder and PC. Server function, set Web to "Yes" (the initial setting is "Yes"). Web page, set Operator or Monitor to "ON" (initial v...

    Publishd:
    Nov, 2017
    ID:
    w0001139
    Category:
    Wiring/Connection/Communciation
  • How can I input a 4-20 mADC signal to the DX100/DX2000?

    Convert the signal to 1-5 VDC with the accessory 250 Ω shunt resistance (for M4 screw terminals, PN 415920)

    Publishd:
    Nov, 2017
    ID:
    w0001140
    Category:
    Specification
  • When input scaling is 0.0–500.0, if the internal data of the measured values is 0.74, what value appears on screen.

    It rounds to one digit lower than the display digit. In this example, 0.74 is rounded to 0.7.

    Publishd:
    Nov, 2016
    ID:
    w0000131
    Category:
    Specification
  • Wants to check the MAC address.

    You can find it in the system information screen or the network information screen. In the operation mode, press FUNC. The FUNC key menu appears. On the FUNC key menu, press the System info soft key or the Network info soft key. The MAC address appears in the system information screen or th...

    Publishd:
    Nov, 2016
    ID:
    w0001124
    Category:
    Setting
  • Error E006 appears on an Advanced Security Function (/AS1 option) model.

    Using the following procedure, reconfirm the registered system administrator information. Enter “root” for the user name, and press DISP/ENTER without entering a password. An error message will appear, followed by a window for entering the password again. Enter the root password, a...

    Publishd:
    Nov, 2016
    ID:
    w0001125
    Category:
    Setting
  • Wants to see report data beyond “Report data 101.”

    The data is saved automatically to the CF card (with the extension, DAR). You can display it on a PC using DAQSTANDARD.

    Publishd:
    Oct, 2016
    ID:
    w0001104
    Category:
    Specification
  • What amperes should be selected for the DXAdvanced series power supply circuit breaker.

    This will be the same as the switch specifications in the operation guide. No /P1 option: Steady-state current rating 1 A or highter, inrush current rating 60 A or higher. With /P1 option (with 24 VDC/AC power supply): Steady-state current rating 3 A or higher, inrush current rating 70 A or highe...

    Publishd:
    Oct, 2016
    ID:
    w0001105
    Category:
    Specification
  • Wants to know how to calculate accuracy when inputting 1–5 VDC and using 0.0–200.0% linear scaling.

    First, calculate the accuracy of voltage measurement at 1–5 V prior to scaling. “rdg” stands for “reading.” When inputting 1 V, it's 0.05% × 1 V + 3 mV = 4 mV (maximum resolution or less, rounded up). In other words, 4 digits. When inputting 5 V, it's 0.05% &t...

    Publishd:
    Oct, 2016
    ID:
    w0001106
    Category:
    Specification
  • Wants to know how many reference junction compensation (RJC) temperature sensors are included.

    One RJC temperature sensor is included with each input terminal block.

    Publishd:
    Oct, 2016
    ID:
    w0001107
    Category:
    Specification
  • Configuring the scan interval

    To set the scan interval, press MENU (to switch to setting mode), hold down FUNC for 3 seconds (to switch to basic setting mode), and select the Menu tab > A/D, Memory. For details, see section 3.1, “Setting the Scan Interval and the Integration Time of the A/D Converter” in the user's...

    Publishd:
    Oct, 2016
    ID:
    w0001108
    Category:
    Setting
  • About the blue and black measured values in the historical trend display.

    The blue values are the maximum/minimum values at the cursor. The black values are the maximum/minimum values of the entire display area.

    Publishd:
    Oct, 2016
    ID:
    w0001109
    Category:
    Specification
  • Whether files undergoing memory sampling can be viewed on a PC.

    During memory sampling, you can use key operations to save data or event data. In Operation mode, press the FUNC key, then press Save display or Save event soft key.

    Publishd:
    Oct, 2016
    ID:
    w0001110
    Category:
    Setting
  • With the /F1 option, wants to know the difference between the “Memory sampling output” and “Output device information on memory stop” settings.

    With memory sampling output, the relay is energized on memory start, and deenergized on memory stop. With the output device information on memory stop setting, the relay is energized when you stop memory sampling. These actions cannot be changed.

    Publishd:
    Oct, 2016
    ID:
    w0001111
    Category:
    Specification
  • Wants to know if they can use differential pressure transmitter output as input, and then perform square root computation.

    Yes. Choose Sqrt for the input range mode.

    Publishd:
    Oct, 2016
    ID:
    w0001112
    Category:
    Setting
  • Unable to input the tag number due to the tag number box not appearing.

    In Basic Setting mode, change the Tag No. Use/Not setting to Use. You can find the procedure in section 5.2, “Displaying Tags or Channel Numbers” of the user's manual (IM 04L41B01-01E).

    Publishd:
    Oct, 2016
    ID:
    w0001113
    Category:
    Setting
  • Configured channel numbers and tag comments to display, but only the channel numbers appear on screen.

    With that setting, both the channel number and tag comments may not be able to fit in the limited display area, and only the channel number is displayed. To display the tag comments, set Tag No. Use/Not to “Not.”

    Publishd:
    Oct, 2016
    ID:
    w0001114
    Category:
    Setting
  • What the values are for 0% and 100% when entering “%” for the position in the Partial Expanded Display settings.

    0% and 100% are the lower and upper limits of span, which is the display range. The boundary value is the position on that range, expressed as a percentage, that divides the compressed and expanded display areas.

    Publishd:
    Oct, 2016
    ID:
    w0001115
    Category:
    Setting
  • Unable to display the current trend and historical trend at the same time when the screen is split in 2.

    The procedure you mentioned can only show historical trends of display data. It won't work with event data.

    Publishd:
    Oct, 2016
    ID:
    w0001116
    Category:
    Setting
  • How to display the scales of both temperature and pressure after setting temperature and pressure to a single group in the display group channel settings.

    Under Setting menu > Measurement channel > Zone scale bar graph, you can set the scale position and number of divisions for each channel. Set the scale positions separately for temperature and pressure. If the scales of multiple channels are set to the same position, the scale of the channel assigne...

    Publishd:
    Oct, 2016
    ID:
    w0001117
    Category:
    Setting
  • Difference in the meaning of “Day” when selecting a report type of Day versus selecting (H+D).

    When you select “Day,” the daily data from each day is added to the report files up to the previous day. In one month, the daily report file is completed. When you select H+D, a report file is created starting from the daily data of the relevant day. The next day, a separate daily report...

    Publishd:
    Oct, 2016
    ID:
    w0001118
    Category:
    Specification
  • In record templates, entered the Tag No. keyword but the tag number doesn't make it into the Excel report file.

    The keyword for tag no. is “ChId.” Please check the keyword again.

    Publishd:
    Oct, 2016
    ID:
    w0001119
    Category:
    Setting
  • Whether able to omit parameters when using record template keywords.

    Parameters are written in the order: report type; report channel number; start date/time; and end date/time. The report type and channel number cannot be omitted, but you can omit the start and end date/time.

    Publishd:
    Oct, 2016
    ID:
    w0001120
    Category:
    Setting
  • Whether able to add multiple record template keywords to a single cell.

    If there are multiple keywords in a single cell, only the first keyword is valid. In your case, you'll need to use separate cells.

    Publishd:
    Oct, 2016
    ID:
    w0001121
    Category:
    Setting
  • How to check the reference temperature of an RTD if the input type is not noted in the /N1 option specifications.

    See the following. Cu10 (GE): at 25°C Cu10 (L&N): at 25°C Cu10 (WEED): at 25°C Cu10 (BAILEY): at 0°C

    Publishd:
    Oct, 2016
    ID:
    w0001122
    Category:
    Specification
  • Whether the size of message and alarm information changes depending on the channel information.

    No, it only varies depending on the number of messages and alarms.

    Publishd:
    Oct, 2016
    ID:
    w0001123
    Category:
    Specification
  • PC operating environment for TUS software (774091)

    The English version of Windows is recommended. Use one of the following operating systems. Windows 7 (32-bit) Windows 7 (64-bit) Windows 8 Windows 2008 Server (32-bit) Windows 2008 Server R2 (64-bit) Windows 2012 Server With Japanese Windows, it may work if you don't use Japanese file n...

    Publishd:
    Aug, 2016
    ID:
    w0000096
    Category:
    Specification
  • Precautions when changing communication on a model with the PROFIBUS-DP communication option (/CP1) to Modbus/TCP communication.

    With the DX1000, you cannot use communication channels. On the DX2000, use communication channels C33–C60 (with the MATH option /M1, you can't use C01–C32).

    Publishd:
    Aug, 2016
    ID:
    w0001102
    Category:
    Specification
  • RTD measurement with the BARD-700

    Yes, you can perform RTD measurements while using the BARD-700 safety barrier. When using the BARD, make it so that the resistance value excluding the BARD is 10Ω or less. Make the resistance the same on each of the 3 wires.

    Publishd:
    Aug, 2016
    ID:
    w0001103
    Category:
    Specification
  • Unable to read data files from the CF card on DAQSTANDARD Viewer

    You may be using an older version of DAQSTANDARD Viewer. Please download the latest version of DAQSTANDARD (DXA120) from our website.

    Publishd:
    Apr, 2016
    ID:
    w0001090
    Category:
    Software
  • Accidentally pressed the stop key after starting recording with a remote signal. Whether recording continues or stops.

    After starting with a remote signal, if you press stop, a confirmation message appears. If you select to stop recording, it stops. If you use key lock on the stop key, recording won't stop even if you accidentally press the stop key during recording.

    Publishd:
    Apr, 2016
    ID:
    w0001091
    Category:
    Setting
  • Wants to obtain the latest version of DAQSTANDARD

    Please download the latest version of DAQSTANDARD (DXA120) from our website. Also, if you lost your license number, please see "Universal keys in case of lost licenses."

    Publishd:
    Apr, 2016
    ID:
    w0001092
    Category:
    Software
  • Wants to know the measured value overrange conditions

    The conditions under which an overrange occurs depends on the input range mode settings. For details, see the note under "Digital Display" in section 1.3 of the User's Manual.

    Publishd:
    Apr, 2016
    ID:
    w0001093
    Category:
    Specification
  • Whether you can expand the trend display area by narrowing the upper and lower limit of the scale when inputting 1-5 VDC and using linear scaling

    Yes. If you narrow the lower and upper-limit values, change the lower and upper-limit values of the span accordingly.

    Publishd:
    Apr, 2016
    ID:
    w0001094
    Category:
    Setting
  • Sets the Modbus client function, but unable to connect to the Modbus server

    Yes, press the DISP/ENTER key, and then select [INFORMATION] > [MODBUS CLIENT]. In order for [MODBUS CLIENT] to appear in the screen menu, you must configure it in the menu customization function. See "Checking the Modbus Operating Status" in section 1.10, "Using the Modbus Client Function" in...

    Publishd:
    Apr, 2016
    ID:
    w0001095
    Category:
    Setting
  • Wants to display the measured temperature 0.5°C higher than actual

    In the input range settings, you can add a specific value in the same manner as a bias setting by selecting a scale. Ex) Set a thermocouple Type-K span of 0-100 DEGC. To display the temperature 0.5DEGC higher, set the following in the input range settings: Mode = Scale; Type = TC; Range = K, Sp...

    Publishd:
    Apr, 2016
    ID:
    w0001096
    Category:
    Setting
  • Wants to change the system administrator password to something other than the default password on an advanced security model (/AS1)

    In the user registration screen, first set the default password. The first time you log in, enter the default password. The message, "Your password has expired. Key in your new password" appears, and you can enter a new password. For more information, see section 2.2, "Logging In and Out" i...

    Publishd:
    Apr, 2016
    ID:
    w0001097
    Category:
    Setting
  • Where USB interface should be inserted

    There are two locations, one on the front and one on the back. On the front, it's inside the front panel door next to the CF card slot. On the back, it's next to the Ethernet interface.

    Publishd:
    Apr, 2016
    ID:
    w0001098
    Category:
    Specification
  • Copying data in internal memory to USB memory

    In the User's Manual, see section 2.12, "Using the USB Flash Memory (/USB1 Option)" and see "Saving the Data" in section 4.8, "Using the Memory Summary."

    Publishd:
    Apr, 2016
    ID:
    w0001099
    Category:
    Setting
  • Ability to save measured data in CSV format

    The DX1000/DX2000 can save measured data in binary format. You can convert saved measured data to ASCII (CSV) format using DAQSTANDARD Viewer.

    Publishd:
    Apr, 2016
    ID:
    w0001100
    Category:
    Specification
  • On the advanced security (/AS1) model, the user locked icon appeared due to password mis-entry during login

    See "Dealing with the Invalid User Status" in section 2.2, "Logging In and Out" in the advanced security function (/AS1) user's manual. If all system administrator passwords are invalidated, please contact your Yokogawa service network.

    Publishd:
    Apr, 2016
    ID:
    w0001101
    Category:
    Setting
  • DAQSTANDARD licenses

    For DAQSTANDARD software licenses, see "Terms and Conditions of the Software License" in the user's manual. The number of licenses is 1. To use multiple PCs, download SMARTDAC+ STANDARD Universal Viewer.

    Publishd:
    May, 2015
    ID:
    w0001085
    Category:
    Software
  • Appearance of error E0211, "Can't write to file." when signing with DAQSTANDARD Viewer

    Check your Windows 7 rights. If you don't have write permission, change your permissions.

    Publishd:
    May, 2015
    ID:
    w0001086
    Category:
    Software
  • Appearance of error E0210, "Not registered user." when signing with DAQSTANDARD Viewer

    It is possible that the user has expired and the password has returned to the default, or some similar problem. See section 2.10, "Signing Batch and Continuous Data" in DXA120 DAQSTANDARD Viewer User's Manual.

    Publishd:
    May, 2015
    ID:
    w0001087
    Category:
    Software
  • Not receiving dedicated DX communication commands when communicating with the Mitsubishi MELSEC

    Some commands are sent in units of words. This always results in an even number of bytes. But if you send the command "FD0,1,1CRLF", for example, that is an odd number of bytes (9), and the 10th byte which follows CRLF will be populated with "junk." This junk byte becomes the 1st byte of the subseq...

    Publishd:
    May, 2015
    ID:
    w0001088
    Category:
    Wiring/Connection/Communciation
  • Support for encrypted USB memory

    No. Encrypted USB memory is not supported.

    Publishd:
    May, 2015
    ID:
    w0001089
    Category:
    Specification
  • Record start/stop using the remote control input option when using the advanced security function

    Yes. For details, see "Event Action Function " and "Remote Control Function " in the user's manual. DX1000 user's manual DX2000 user's manual

    Publishd:
    Jul, 2014
    ID:
    w0001022
    Category:
    Specification
  • Ethernet communication not working

    Connect the PC and DX via a hub. If the connection is working, the Link LED on the PC, hub, and DX should all light. Check the PC's IP address and subnet mask using the ipconfig command. Check the DX's IP address and subnet mask in the Network screen by pressing the FUNC key. Confirm that the PC ...

    Publishd:
    Jul, 2014
    ID:
    w0001054
    Category:
    Wiring/Connection/Communciation
  • Writing MELSEC data during Modbus/TCP communications

    Yes. However, the DX2000 requires the MATH function option (suffix code: /M1). The procedure is as follows. Enable the DX2000's Modbus server function. Register communication input channels (C01 or C02) in a formula and start the MATH function. Next, assign MATH channels to groups, and set them ...

    Publishd:
    Jul, 2014
    ID:
    w0001026
    Category:
    Wiring/Connection/Communciation
  • Unable to open files in DAQSTANDARD

    You may have an older version of DAQSTANDARD. Consider upgrading to the latest version. You can download the latest version by using the following link. DAQSTANDARD download page

    Publishd:
    Jul, 2014
    ID:
    w0001046
    Category:
    Software
  • Terminal wiring and terminal screw diameter

    For terminal wiring, see the section on grounding and wiring in the operation guide. DX1000 Operation Guide DX2000 Operation Guide The diameter of all the screws is 4 mm. DX1000 Operation Guide DX2000 Operation Guide

    Publishd:
    Jul, 2014
    ID:
    w0001055
    Category:
    Specification
  • Viewing data that was recorded onto the CF card on a PC

    Yes, by using the accessory software DAQSTANDARD (model DXA120). See chapter 3, "Viewing Recorded Data " in the operation guide. For further details, see the viewer's user's manual. DXA120 user's manual

    Publishd:
    Jul, 2014
    ID:
    w0001056
    Category:
    Specification
  • Creation of report files upon power failures

    No, but the creation of the report restarts after recovery from the power failure. The operation is different depending on whether the recovery time is before or after the report creation time has elapsed. For details, see "Special Data Handling " in section 1.8, "Computation and Re...

    Publishd:
    Jul, 2014
    ID:
    w0001058
    Category:
    Specification
  • Changing the measured data interval (recording interval)

    The DX has two kinds of files, DISP (display data) and EVENT (event data). You can tell which type the data is by looking inside the blue band in the upper part of the screen. In the case of (2) DISP (display data), the sampling interval changes in sync with the trend update interval [/div]. For de...

    Publishd:
    Jul, 2014
    ID:
    w0001062
    Category:
    Setting
  • Comparing settings data

    Yes, by printing the settings from the DAQSTANDARD (accessory software) hardware configurator. For details, see section 1.5, "Printing Setup Data" in the DAQSTANDARD hardware configurator user's manual. DAQSTANDARD user's manual

    Publishd:
    Jul, 2014
    ID:
    w0001064
    Category:
    Specification
  • Loss of DAQSTANDARD license number

    We have a special license number available in case of lost licenses. See the following URL.   Universal keys in case of lost licenses

    Publishd:
    Jul, 2014
    ID:
    w0001066
    Category:
    Specification
  • Logon test method for a DX connected to a PC via Ethernet

    You can use the Windows telnet command. Click Start > All programs > Accessories > Command Prompt. At the command prompt, type "telnet. " Given a DX IP address of "10.1.32.130," type the following (the port number is 34260). open 10.1.32.130 34260 When E1 402...

    Publishd:
    Jul, 2014
    ID:
    w0001077
    Category:
    Specification
  • Error E282, "An error occurred with the FTP controller connection."

    Communication could not be established with the FTP server. You can check details on codes in the FTP log screen. In the FTP log screen, while DX1000/DX2000 is running, press the DISP/ENTER key. In the operation menu on the left side of the screen, move to "Log, " then use the right ar...

    Publishd:
    Jul, 2014
    ID:
    w0001080
    Category:
    Setting
  • Error E201, "Insufficient free space on media"

    Yes, turn ON the media FIFO function. If the CF card does not have sufficient free space, or if there are 1000 files or more, old files will be overwritten with new ones. For details on the media FIFO function, see Explanation of functions: Data Saving Function in the user's manual. DX1000/DX1...

    Publishd:
    Jul, 2014
    ID:
    w0001081
    Category:
    Specification
  • Clearing the trigger mark in DAQSTANDARD Viewer

    Yes. Select a trigger mark with the cursor, then click Edit > Delete Mark. For details, see section 2.2, "Displaying Waveforms " in the DAQSTANDARD Viewer user's manual. User's manual download

    Publishd:
    Jul, 2014
    ID:
    w0001082
    Category:
    Specification
  • Displaying "Legend" in the DAQSTANDARD Viewer's trend graph

    No, but you can with the SMARTDAC+ universal viewer. The SMARTDAC+ universal viewer supports data from the DX and MV series, and is available for free download from our Web server. SMARTDAC+ Standard Universal Viewer download page

    Publishd:
    Jul, 2014
    ID:
    w0001083
    Category:
    Specification
  • Support for the RoHS directive

    Please see the document below. You can confirm the compliance status of recorders and controllers with the RoHS directive. RoHS directive compliance status

    Publishd:
    Jul, 2014
    ID:
    w0000046
    Category:
    Other
  • FDA 21 CFR Part 11 support

    The DX series does with the advanced security option (suffix code /AS1).

    Publishd:
    Jul, 2014
    ID:
    w0001001
    Category:
    Model selection
  • Advanced security and Multibatch functions

    Yes, you can use the Advanced security option (suffix code: /AS1) and the Multibatch function option (suffix code: /BT2) together.

    Publishd:
    Jul, 2014
    ID:
    w0001002
    Category:
    Specification
  • PROFIBUS-DS communication interface and Advanced security function

    Yes, you can use the PROFIBUS-DP communication option (suffix code: /CP1) and the Advanced security function option (suffix code: /AS1) together.

    Publishd:
    Jul, 2014
    ID:
    w0001003
    Category:
    Specification
  • Pulse input and MATH function

    The pulse input option (suffix code: /PM1) includes the MATH function option (suffix code: /M1). By adding the pulse input option, you can also use the MATH function. It is not necessary to add both options together.

    Publishd:
    Jul, 2014
    ID:
    w0001004
    Category:
    Specification
  • Replacing µR series chart recorders

    Yes, the µR series panel cutout dimensions are the same as those of the DX2000, so you can replace them easily.

    Publishd:
    Jul, 2014
    ID:
    w0001005
    Category:
    Model selection
  • DC24V power supply

    Yes, if you add the 24 VQAC/DC power option (suffix code: /P1) you can run it on DC24 V power.

    Publishd:
    Jul, 2014
    ID:
    w0001006
    Category:
    Specification
  • Changing SNTP settings during data acquisition

    No. When changing settings, data acquisition stops.

    Publishd:
    Jul, 2014
    ID:
    w0001007
    Category:
    Setting
  • Time setting via external contact input

    No, the time cannot be specified.

    Publishd:
    Jul, 2014
    ID:
    w0001008
    Category:
    Specification
  • Remote control input option

    There are 8 input terminals. They are of the non-voltage contact and open collector types. The signal types are level and edge (250 ms or higher).

    Publishd:
    Jul, 2014
    ID:
    w0001009
    Category:
    Specification
  • Maximum number of pulses that can be acquired when using pulse input

    The pulse input option (suffix code: /PM1) provides 3 dedicated pulse inputs plus 5 remote control common terminals for a total of 8 acquisitions. Integral values can also be calculated on individual MATH channels.

    Publishd:
    Jul, 2014
    ID:
    w0001010
    Category:
    Specification
  • Changing display languages

    Yes. You can change it to Japanese, English, German, French, or Chinese.

    Publishd:
    Jul, 2014
    ID:
    w0001011
    Category:
    Specification
  • Saving data upon power failure

    Yes, the internal memory is non-volatile memory, therefore the data just before the power failure is saved. When the instrument recovers from a power failure, data acquired just before the power failure are saved as files to internal memory. Also, data acquisition is restarted.

    Publishd:
    Jul, 2014
    ID:
    w0001012
    Category:
    Specification
  • Display screen customization

    Yes, with R3 (release 3). With DAQStudio (DXA250) screen editing software, you can easily create custom display screens.

    Publishd:
    Jul, 2014
    ID:
    w0001013
    Category:
    Specification
  • Loading MW100 measured data on DXAdvanced.

    Yes. However, the DX1000 requires the MATH function option (suffix code: /M1), and the DX2000 requires the expansion channel function option (suffix code: /MC1) or the MATH function option (suffix code: /M1).

    Publishd:
    Jul, 2014
    ID:
    w0001014
    Category:
    Wiring/Connection/Communciation
  • Increasing the number of inputs

    You cannot increase the number of input terminals. However, you can load data acquired on other instruments such as the MW100 onto the DX via communications.

    Publishd:
    Jul, 2014
    ID:
    w0001015
    Category:
    Specification
  • Modbus communications

    Yes. Adding the RS-422/485 communication interface option (suffix code: /C3) makes the DX a Modbus/RTU subordinate, enabling communications with Modbus master devices. Note that making the DX the Modbus master requires the MATH function option (suffix code: /M1).

    Publishd:
    Jul, 2014
    ID:
    w0001016
    Category:
    Wiring/Connection/Communciation
  • Anti data tampering

    Saving measured data in binary format prevents it from being tampered with.

    Publishd:
    Jul, 2014
    ID:
    w0001019
    Category:
    Specification
  • Restrictions on PROFIBUS communications when using the advanced security function

    Even when the advanced security option (suffix code: /AS1) is running, PROFIBUS communication can proceed without restriction. No login via ID or password is required.

    Publishd:
    Jul, 2014
    ID:
    w0001020
    Category:
    Specification
  • Restrictions on saving to USB memory when using the advanced security function.

    Yes. With the user login function, you can set operations for each user.

    Publishd:
    Jul, 2014
    ID:
    w0001021
    Category:
    Specification
  • Changing group and alarm settings when using the Advanced security and Multibatch functions

    When using the Multibatch function option (suffix code: /BT2), you cannot change the range even if group 1 is recording, regardless of any other options. However, you can edit alarm settings during recording.

    Publishd:
    Jul, 2014
    ID:
    w0001023
    Category:
    Setting
  • Ethernet connection between the DX2000 and the MW100

    The DX2000 (client) requires the expansion channel function option (suffix code: /MC1) or the MATH function option (suffix code: /M1). The MW100 (server) does not require the MATH function. Both serial and Ethernet can be used together. In that case, the DX2000 will require the RS-422A/485 communica...

    Publishd:
    Jul, 2014
    ID:
    w0001024
    Category:
    Wiring/Connection/Communciation
  • Modbus/TCP communications between two DXAdvanced units

    Yes. However, this requires the MATH function option (suffix code: /M1) on the unit that is displaying (recording) data.

    Publishd:
    Jul, 2014
    ID:
    w0001025
    Category:
    Wiring/Connection/Communciation
  • FTP transfer during network interference

    Unsent files will be sent upon the next FTP transfer.

    Publishd:
    Jul, 2014
    ID:
    w0001027
    Category:
    Wiring/Connection/Communciation
  • Simultaneous communication with multiple PCs

    Yes. The communication port used by Modbus is different than the one that can acquire instantaneous values, so simultaneous communication is possible.

    Publishd:
    Jul, 2014
    ID:
    w0001028
    Category:
    Wiring/Connection/Communciation
  • Custom screen display of data acquired by Modbus communication

    Yes. You can also display communication input channels, MATH channels, and expansion channels.

    Publishd:
    Jul, 2014
    ID:
    w0001029
    Category:
    Wiring/Connection/Communciation
  • FTP transfer time difference setting

    When the FTP server gets file transfer requests from multiple DXs or other instruments, it cannot handle all of those requests at once. This setting delays the transfer times to avoid errors.

    Publishd:
    Jul, 2014
    ID:
    w0001030
    Category:
    Specification
  • Life span of the built-in battery

    For best results, we recommend replacing the built-in lithium battery every 10 years. Note that battery replacement should be performed by a trained Yokogawa technician. Please contact your nearest dealer or Yokogawa representative.

    Publishd:
    Jul, 2014
    ID:
    w0001031
    Category:
    Specification
  • Calibration scheduling function

    This function allows you to control periodic entering of settings for the calibration function option (suffix code: /CC1). When you enable the calibration function, a screen appears that displays the calibration due date in advance (the calibration notification screen).

    Publishd:
    Jul, 2014
    ID:
    w0001032
    Category:
    Specification
  • Start/stop of recording per external contact input and auto saving of data

    Yes, this is possible by adding the remote control function option (suffix code: /R1) and using the event action function. Note that you can also automatically save data to the CF card while recording is stopped.

    Publishd:
    Jul, 2014
    ID:
    w0001033
    Category:
    Specification
  • Loading of data from internal memory onto a PC

    Yes, when the FTP server is ON.

    Publishd:
    Jul, 2014
    ID:
    w0001034
    Category:
    Specification
  • Auto saving data onto USB memory

    No. Due to reliability issues with recording media, you can only save data automatically onto the CF card. The USB memory is only intended for temporarily transferring data files.

    Publishd:
    Jul, 2014
    ID:
    w0001035
    Category:
    Specification
  • Supported CF cards

    Yes, the following: 772093: 512 MB 772094: 1 GB 772095: 2 GB

    Publishd:
    Jul, 2014
    ID:
    w0001036
    Category:
    Specification
  • Handling a full CF card

    Insert the CF card into a PC and transfer the card's data files to the PC's hard disk. You can view the files saved on the PC by using DAQSTANDARD, the program that comes with the DX.

    Publishd:
    Jul, 2014
    ID:
    w0001037
    Category:
    Specification
  • CF card type

    It is a type 1 card.

    Publishd:
    Jul, 2014
    ID:
    w0001038
    Category:
    Specification
  • Accessories that come standard

    Only the CF card comes standard. The card adapter is not included. The CF card can be inserted into the main unit directly.

    Publishd:
    Jul, 2014
    ID:
    w0001039
    Category:
    Other
  • Limit on the number of files saved when using the media FIFO function

    Yes. When using the media FIFO function, the maximum number of files that can be created in a single directory is 1000. If that number is exceeded, old files will be overwritten with new ones.

    Publishd:
    Jul, 2014
    ID:
    w0001040
    Category:
    Specification
  • Display when the CF card starts running out of free space

    Yes. When there is 10% of free space or less, the CF card icon on the upper part of the display turns from green to red.

    Publishd:
    Jul, 2014
    ID:
    w0001041
    Category:
    Specification
  • Error when inserting the CF card

    There may be a problem with the CF card. Try the following. Eject, then reinsert the CF card. Replace the card. Reformat the card on the DX.

    Publishd:
    Jul, 2014
    ID:
    w0001042
    Category:
    Specification
  • Ejecting the CF card during recording

    Yes, if the CF card is not being accessed. To eject it, press the Func key, then select Eject media > CF.

    Publishd:
    Jul, 2014
    ID:
    w0001043
    Category:
    Specification
  • Saving data and FTP transfer when there is insufficient space on the CF card

    When the CF card becomes full, files continue to be created in the internal memory. The files created in internal memory are also transferred via FTP.

    Publishd:
    Jul, 2014
    ID:
    w0001044
    Category:
    Specification
  • Saving report data on the CF card

    As with recorded data, files are saved to the CF card after being created in internal memory if the auto save function is ON. Hourly reports are saved every hour, and daily reports are saved every day.

    Publishd:
    Jul, 2014
    ID:
    w0001045
    Category:
    Specification
  • DAQSTANDARD functions

    DAQSTANDARD is composed of 2 programs. Viewer: You can display measured data as trends or numeric values, and print them. Plus, binary data files can be exported to ASCII or Excel format. Hardware configurator: You can create settings files for a recorder, send data to the recorder, and save dat...

    Publishd:
    Jul, 2014
    ID:
    w0001047
    Category:
    Software
  • Data display and analysis

    Yes, by using the DAQSTANDARD software that comes standard.

    Publishd:
    Jul, 2014
    ID:
    w0001048
    Category:
    Software
  • Creating custom display screens

    To create them on a PC you will need the DXA170 DAQStudio custom display builder software, sold separately. You can create them by using keys on the main unit, but it is much easier by using DAQStudio. DAQStudio Product Page

    Publishd:
    Jul, 2014
    ID:
    w0001049
    Category:
    Specification
  • Successor to the DX100P/DX200P

    The successor to the DX100P/DX200P would be the DX1000/DX2000 with the advanced security function option (suffix code: AS1). Data files recorded on the DX100P/DX200P can be opened using DAQSTANDARD (DXA120), which comes with the DX1000/DX2000.

    Publishd:
    Jul, 2014
    ID:
    w0001050
    Category:
    Model selection
  • User's manuals included upon delivery

    Three printed manuals are included, the "Operation Guide, " "DXA120 Installing DAQSTANDARD, " and "Keeping the Product Clean. " PDF user's manuals are also included on CD-ROM.

    Publishd:
    Jul, 2014
    ID:
    w0001051
    Category:
    Other
  • Inclusion of a warranty card

    The warranty is handled with the DX serial number, so a printed warranty card is not included. The warranty period is one year from date of purchase.

    Publishd:
    Jul, 2014
    ID:
    w0001052
    Category:
    Other
  • Validated compact flash cards (CF cards)

    Please see the following. List of validated compact flash (CF cards)

    Publishd:
    Jul, 2014
    ID:
    w0001053
    Category:
    Specification
  • PC acquisition of data files from the CF card or internal memory

    Yes, by using 3rd party FTP client software.

    Publishd:
    Jul, 2014
    ID:
    w0001057
    Category:
    Wiring/Connection/Communciation
  • Supported e-mail authorization method

    They are "AUTH PLAIN " and "AUTH CRAM-MD5. " AUTH-LOGIN is not supported.

    Publishd:
    Jul, 2014
    ID:
    w0001059
    Category:
    Specification
  • Adjusting the main unit time during recording

    Yes, you can adjust it directly on the main unit even during recording. If you adjust the time during recording, the acquired data is time-stamped with the new time.

    Publishd:
    Jul, 2014
    ID:
    w0001060
    Category:
    Specification
  • Data list not appearing in the web server screen

    The following conditions must be met in order for the data list to appear in the web server screen. The FTP server must be turned ON. Login via communication must not be set. Note that the data list never displays on models with the advanced security function option (suffix code: /AS1).

    Publishd:
    Jul, 2014
    ID:
    w0001061
    Category:
    Specification
  • Multiple mis-entries of user ID and password

    You can set the number of password retries to 3, 5, or unlimited (default is 3). When the number of retries is exceeded, the unit is locked. Service is required to clear the lock, so please manage user IDs and passwords with care. Note that if 1 unit has multiple administrators configured, we recomm...

    Publishd:
    Jul, 2014
    ID:
    w0001063
    Category:
    Specification
  • Automatic conversion of recorded data files to Excel

    Yes. With DAQManager, you can specify the folder containing the recorded data files as the auto-import destination, then set it for automatic conversion to Excel. Note that this is not supported with addition of the advanced security option (suffix code /AS1) to the DX.

    Publishd:
    Jul, 2014
    ID:
    w0001065
    Category:
    Specification
  • Receiving settings from a DX with the advanced security function

    You can receive settings by using the user management "Monitor" function specific to advanced security. The Monitor function enables up to 2 connections. When 2 Monitor functions are connected on the network, you can receive settings.

    Publishd:
    Jul, 2014
    ID:
    w0001076
    Category:
    Specification
  • Time setting on the DX connected from DAQEXPLORER

    Yes. It is set manually, under the Set Time.

    Publishd:
    Jul, 2014
    ID:
    w0013030
    Category:
    Setting
  • Handling PC shutdowns while DAQLOGGER is running

    If DAQLOGGER stops while communicating with the DX1000, depending on the timing, the DX1000 might enter a response wait state and be unable to receive any other commands. With apologies, please turn OFF the DX1000 and restart DAQLOGGER.

    Publishd:
    Jul, 2014
    ID:
    w0013059
    Category:
    Wiring/Connection/Communciation
  • Monitor switching when changing alarm values on DAQLOGGER

    Unfortunately you will need to stop DAQLOGGER. When you rescan, the changed alarm values will apply.

    Publishd:
    Jul, 2014
    ID:
    w0013060
    Category:
    Specification
  • Checking account information on a DX with the advanced security function

    No.

    Publishd:
    Apr, 2014
    ID:
    w0001078
    Category:
    Model selection
  • Precautions when using "All settings" on a DX with the advanced security function

    Yes, note that the passwords in the account information will be initialized.

    Publishd:
    Apr, 2014
    ID:
    w0001079
    Category:
    Specification
  • Trend display when setting the trip line in DAQSTANDARD Viewer

    If assigned to multiple channels, display the trip line of the active waveform. Click the color bar on the left side of the viewer, and activate the waveform of the channel on which you set the trip line.

    Publishd:
    Apr, 2014
    ID:
    w0001084
    Category:
    Specification